77
────────────────────────────────────────────────────
5.3 ST ATUS Settings (REC)
────────────────────────────────────────────────────
NOTE
5.3.10 Setting the Interpolation (dot-line, X-Y only)
Function
display Meaning
:
Linear interpolation is not performed.
:
Linear interpolation is performed.
5.3.11 Setting the Display Clear Function (X-Y only)
Function
display Meaning
:
Display clear is disabled.
:
Display clear is enabled.
・
File names consist of 8 characters (or 4 double-byte characters). With auto-
save, since numbers are attached to the end of file names, long file names
are truncated when this number is appended to the end of the file name.
・
With auto-save enabled, channel selection is disabled. Data for all channels
is saved.
・
For details on file name of auto saving, refer to Section 14.7.
・
When the recording length is "CONT.", data are not stored.
・
For details on connected SCSI device, refer to Section 14.4.
・
When cursor A and B are enabled, partial saving is executed.
・
If additional recording is enabled, only newly acquired waveforms are saved.
Even when A and B cursors appear on the screen, no partial save is
performed.
・
During automatic storage, if the measurement is interrupted, waveforms
taken prior to the interruption are stored automatically.
・
The directory and the number of files that can be stored in the directory are
limited. For details, see Section 14.7.
・
When "COMMUNI" (communications) is selected as Media, Storage Method
and Type are not displayed.
・
Interpolation can be set when set the X-Y graph format.
・
This setting determines whether the input waveform (sampling data) is to be
displayed and printed as a series of dots or a line using linear interpolation.
Procedure
Screen: STATUS
1. Press the STATUS key to display the STATUS screen.
2. Move the flashing cursor to the dot-line item.
3. Use the function keys to make a setting.
・
It sets whether clear the waveform or not when pressing the START key to
start measurement and the previous waveform is left.
・
If the waveform is not cleared, overlay is performed.
Procedure
Screen: STATUS, Waveform display
1. Use the Menu keys to display the desired screen.
2. Move the flashing cursor to the display clear item.
3. Use the function keys to make a setting.