8.4 STATUS2 Settings (FFT)
────────────────────────────────────────────────────
Input
Trigger conditions
fulfilled
Waveform
capture
FFT Averaging Display End
Press the STOP key
If the trigger condition does not occur within the specified period,
waveform data input begins anyway.
8.4.2 Setting the Interpolation (dot-line)
Function
display
Meaning
:
Linear interpolation is not performed.
:
Linear interpolation is performed.
(3) Trigger mode: AUTO
(Time axis waveform)
When the START key is pressed, data are captured even if trigger conditions
are not fulfilled after a certain interval. If averaging is applied to
unsynchronized input signals, the result will be meaningless.
(Frequency axis waveform)
1. After the START key was pressed, data are captured whenever the trigger
conditions are fulfilled, and averaging is carried out until the specified
averaging count. The averaging result is shown on the display.
2. When the specified number of data has been captured, measurement stops
automatically.
3. If the measurement was stopped prematurely with the STOP key, the
averaging result up to that point is displayed.
This setting determines whether the input waveform (sampling data) is to be
displayed and printed as a series of dots or a line using linear interpolation.
Procedure
Screen: STATUS2
1. Press the STATUS key to display the STATUS2 screen.
2. Move the flashing cursor to the dot-line item.
3. Use the function keys to make a setting.