2.7 9322 DIFFERENTIAL PROBE Connection
────────────────────────────────────────────────────
DANGER
When using grabber clips, the 9322's maximum rated voltage to
earth is 1500 VAC/DC; when using alligator clips, it is 1000 VAC/DC.
To avoid electrical shock and possible damage to the unit, never
apply voltages greater than these limits between the input channel
terminals and chassis, or across the inputs of two 9322s.
Maximum input voltage is 1000 VAC/2000 VDC. Do not measure
voltage in excess of these limitations, as doing so may damage the
unit or cause an accident that might result in injury or death.
2.7 9322 DIFFERENTIAL PROBE Connection
9322 is a differential probe that connects to input of 8841/42 MEMORY
HiCORDER input unit. For more details, refer to its instruction manual.