368
────────────────────────────────────────────────────
16.2 Calcu lating Waveform Data
────────────────────────────────────────────────────
16.2.5 Setting the Display Scale and Number of Moved Points
s1425-1
Function
display Meaning
#g-eq7
:
Set the scale and floating decimal point.
Function
display
Meaning
#g-sc1, 2
:
Set display scale automatically.
:
Set display scale manually.
:
End
Display scale can be set automatically or manually.
The channel selected for recording is automatically set to variable display.
(See Section 9.7)
If MOV or SLI was used, the number of moved points must be specified.
Procedure
Screen: STATUS4
1. Press the STATUS key to display the STATUS4
screen.
2. Move the flashing cursor to
SETTINGS
.
3. Select the
SETTINGS
function key.
Setting the display scale
1. Move the cursor to Scale.
2. Make the setting with the function keys.
Automatic setting
After calculation, the upper and lower limit is determined from the result,
and the variable display settings are made accordingly.
Depending on the type of calculation, automatically display scale setting may
not be satisfactory. In such a case, use the manual setting procedure.
Manual setting
Use the variable display setting function on the VARIABLE screen to set the
upper and lower limit. (See Section 9.7.)
Calculation result with overflows
The values shown using cursors A and B and the printed values obtained
when the printer recording type is set to "numerical value" are not accurate.
If the display scale is set to "
AUTO
," the waveform is shown at the top or
bottom of the screen. This indicates that the calculation result has
overflowed.