206
────────────────────────────────────────────────────
9.14 8940 F/V UNIT
────────────────────────────────────────────────────
Flashing cursor
Flashing cursor
Function
display Meaning
:
Pull-up resistance is enabled (ON).
(for connection to open collector output)
:
Pull-up resistance is disabled (OFF).
Function
display
Meaning
:
Increases in number, large step
:
Increases in number, small step
:
Decreases in number, small step
:
Decreases in number, large step
(3) Set the pull-up resistance
Set the pull-up resistance on or off. Pull-up resistance
is used when connecting to an open collector output
signal. For normal measurements, disable the pull-up
resistance (set to OFF).
1. Move the flashing cursor to the position shown in the
figure on the left.
2. Use the function keys to set.
(4) Set the threshold level.
1. Move the flashing cursor to the position shown in the
figure on the left.
2. Use the function keys, or the Jog/Shuttle control to
set the threshold value.