EasyManua.ls Logo

Hioki 8841 - Page 454

Hioki 8841
494 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
APPENDIX 10
────────────────────────────────────────────────────
Appendix 3 Reference
────────────────────────────────────────────────────
Appendix 3.4 Recorder Function
1234
・・・・・・・
・・・・・・・
100
1 DIV
Maximum voltage
Minimum voltage
One piece of sample data
Data obtained in the
set sampling period
1234
・・・・・・・
・・・・・・・
100
1 DIV
Maximum voltag
e
Minimum voltag
e
One piece of sample data
Set a short sampling period
One division is equal to 100 samples.
One piece of sample data collected using the recorder function contains the
maximum and minimum voltage obtained in the set sampling period.
Therefore, this data is of a certain width.
When a short sampling period is set and the input waveform changes
slightly, a sudden disturbance such as noise will increase the difference the
between the maximum and minimum values. To eliminate this phenomenon,
set a long sampling period.

Table of Contents

Related product manuals