APPENDIX 15
────────────────────────────────────────────────────
Appendix 3 Reference
────────────────────────────────────────────────────
X-Y waveform
Shaded area is calculated
No enclosed range,
Line area is calculated
NOTE
(14) X-Y area value
・Calculates the area (V
2
) after X-Y plotting.
・The waveform is plotted on the X-Y screen, and the area enclosed by the
plot lines is calculated.
・In single, dual, or quad screen, the A・B cursors (vertical, trace) can be used
to specify the range (see Section 11.2.) for X-Y plotting and area calculation.
・On the X-Y screen of the memory recorder function, it is not possible to
specify the range with the A・B cursors.
・ Depending on the signal waveform, values for parameters (8), (9), (10), and
(11) may not be displayed.
・ When the scaling function is used, scaling is first applied to waveform data,
and then the parameters are calculated. The parameter unit is determined by
the scaling unit (see Section 9.8).