59
────────────────────────────────────────────────────
4.6 Auto Range Function
────────────────────────────────────────────────────
NOTE
Because the auto-range function performs automatic setting for the input
signal present at the time the function is executed, input a signal before
executing the function.
If for the input signal for this channel there is only a small difference
between the maximum value and the minimum value in the range of highest
sensitivity (5 mV/DIV), the setting is made by taking the next higher
channel.
If the range cannot be determined, for all channels for which the waveform
display is on. A warning message appears, and measurement is abandoned.
When the auto range function is activated by pressing the
AUTO
key, a
trigger output signal is generated. This should be taken into consideration
when using both the trigger output and the auto range function.
The auto-range function does not operate on channels for which the
CHARGE or PREAMP measurement mode is selected on the 8947
CHARGE UNIT.