4535 611 98931 iE33 Service Manual Page 254
CSIP Level 1 Troubleshooting: System Software Tools
run and 2) that the Host Motherboard and Philips Software Application is functional enough
to execute the Test and Utilities application.
• Narrative - A narrative of how the test executes, values it may use, special factors with the
test, and information that helps a technician understand how to troubleshoot a failed compo-
nent.
• Subjective tests - (testing FRUs such as the control panel, main display monitor, and speakers)
These tests are not automated pass/fail tests. The Test and Utilities display provides access to
the available test, but you determine whether the results are correct.
NOTE save Tech Admin data, clear all of the log files, and transfer system files Using Data
Transfer enables you to clear the log files without erasing the service logs. If you clear
the logs using MTE, the service logs are erased also.
Help
The Technical Administration (Tech Admin) Help utility (not to be confused with the clinical user
information Help) supports users of the Tech Admin functions. Tech Admin Help provides an
overview of the Tech Admin Tests and Utilities functions and instructions about how to use the
functions.
Tech Admin Help is also available on an AUSSI CD. You can use the CD to view the Tech Admin
Help information on a personal computer.