EasyManua.ls Logo

Philips iE33 - Page 91

Philips iE33
480 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4535 611 98931 iE33 Service Manual Page 91
CSIP Level 1 Theory of Operation: Physical Structure
Neo Analog Interface Module (NAIM
Monitors system temperature, voltage, and beamformer pulser power consumption
Controls Scanhead Select interface (transducer select, personality retrieval, and connectivity
status)
Transducer personality data identifies which transducer(s) are connected to the system
Generates TGC1 and TGC2 waveforms
Supplies TGC voltages to the Channel Boards for non-Matrix transducers or to the beam-
former resident in Matrix transducers
Pulser regulators (±20-80 Vdc, ±60-120 Vdc) and LV CW regulator with disable support
HV MUX current limited source voltages
Regulates acquisition analog supply
Pulser power monitor source
High Speed Serial Interface to Scanhead Select
TGC and HVMUX data demultiplexing and signal conversion
Contains 320 MHz master oscillator and generates clock signals for all Acquisition hardware
(64 MHz, 40 MHz, 32 MHz and Link_Clk) distributed as differential PECL. Clock_Sync is dis-
tributed as LVCMOS
A/D data acquisition block for measuring transducer mounted sensors, acquisition system
voltages, pulser supply voltage and current and system temperature transducers
Power supply regulators and current limits for all voltages supplied to the transducer array
connectors (±12 Vdc, +5 Vdc, +REF, and ±HV MUX bias)
Local regulation of ±15 V to ±12 V for transducer use
Local generation of +10 V reference for TGC waveform D/As and transducer use

Table of Contents

Other manuals for Philips iE33

Related product manuals