Section
3: Instrument description 2470 High Voltage SourceMeter Instrument
Reference Manual
3-50 2470-901-01 Rev. A / May 2019
Reset the instrument
You can reset many of the instrument settings to their default values. For detail on what gets reset,
see Reset default values (on page 5-28
). Default values are also listed in the command descriptions.
If you are connected to a TSP-Link system, resetting the instrument resets all TSP-Link enabled
instruments on the TSP-Link system.
Using the front panel:
1. Press MENU.
2. Under System, select Info/Manage.
3. Select System Reset.
4. The commands are reset and a confirmation message is displayed.
Using SCPI commands:
Send the command:
*RST
Using TSP commands:
Send the command:
reset()
If the instrument is connected to a TSP-Link system and you are using TSP commands, you can
reset only the local instrument by sending localnode.reset() instead of reset().
Using the event log
The event log records events, which can be errors, warnings, and information reported by the
instrument. Through the Event Log menu, you can view these events. You can also specify which
events are shown in the event log, which ones are logged, and which ones generate popup
messages.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals