2: Installation 2470 High Voltage SourceMeter Instrument
Reference Manual
2-40 2470-901-01 Rev. A / May 2019
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
System information
You can get the serial number, firmware build, detected line frequency, calibration verify date,
calibration adjust date, and calibration adjust count information from the instrument.
To view the version and serial number information from the front panel:
1. Press the MENU key.
2. Under System, select Info/Manage.
The firmware version and serial number are displayed at the top of the screen.
To view the calibration information from the front panel:
1. Press the MENU key.
2. Under System, select Calibration.
The adjust date, adjust count, and calibration date are displayed.
To view the line frequency information from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Scroll down to display the line frequency.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals