14: TSP command reference 2470 High Voltage SourceMeter Instrument
14-274 2470-901-01 Rev. A / May 2019
Example
LinkMaster = tsplink.master
Store the TSP-Link master node number in a
variable called
.
Also see
tsplink.initialize() (on page 14-270)
tsplink.node
This attribute defines the node number.
Type TSP-Link accessible Affected by Where saved Default value
Usage
nodeNumber = tsplink.node
tsplink.node = nodeNumber
The node number of the instrument or enclosure (1 to 63)
Details
This command sets the TSP-Link node number and saves the value in nonvolatile memory.
Changes to the node number do not take effect until tsplink.reset() from an earlier TSP-Link
instrument or tsplink.initialize() is executed on any node in the system.
Each node connected to the TSP-Link system must be assigned a different node number.
Example
Sets the TSP-Link node for this instrument to number 3.
Also see
tsplink.initialize() (on page 14-270)
tsplink.state (on page 14-275)
tsplink.readport()
This function reads the TSP-Link trigger lines as a digital I/O port.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
data = tsplink.readport()
Numeric value that indicates which lines are set