14: TSP command reference 2470 High Voltage SourceMeter Instrument
14-216 2470-901-01 Rev. A / May 2019
Example
trigger.model.setblock(1, trigger.BLOCK_BUFFER_CLEAR)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(3, trigger.BLOCK_DELAY_CONSTANT, 0.1)
trigger.model.setblock(4, trigger.BLOCK_BRANCH_COUNTER, 10, 2)
trigger.model.initiate()
delay(1)
print(trigger.model.getbranchcount(4))
Reset trigger model settings.
Clear defbuffer1 at the beginning of the trigger model.
Loop and make five readings.
Delay 0.1 s.
Loop ten more times back to block 2.
Send the count command to check which count has been completed for block 4.
At end of execution, 10 readings are stored in
.
Also see
trigger.model.setblock() — trigger.BLOCK_BRANCH_COUNTER (on page 14-227)
trigger.model.initiate()
This function starts the trigger model.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
trigger.model.initiate()
Also see
Trigger model (on page 8-27)
trigger.model.abort() (on page 14-214)