Section
14: TSP command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
14-216 2470-901-01 Rev. A / May 2019
Example
reset()
trigger.model.setblock(1, trigger.BLOCK_BUFFER_CLEAR)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(2, trigger.BLOCK_MEASURE)
trigger.model.setblock(3, trigger.BLOCK_DELAY_CONSTANT, 0.1)
trigger.model.setblock(4, trigger.BLOCK_BRANCH_COUNTER, 10, 2)
trigger.model.initiate()
delay(1)
print(trigger.model.getbranchcount(4))
waitcomplete()
Reset trigger model settings.
Clear defbuffer1 at the beginning of the trigger model.
Loop and make five readings.
Delay 0.1 s.
Loop ten more times back to block 2.
Send the count command to check which count has been completed for block 4.
At end of execution, 10 readings are stored in
defbuffer1
.
Also see
trigger.model.setblock() — trigger.BLOCK_BRANCH_COUNTER (on page 14-227)
trigger.model.initiate()
This function starts the trigger model.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
trigger.model.initiate()
Also see
Trigger model (on page 8-27)
trigger.model.abort() (on page 14-214)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals