12: SCPI command reference 2470 High Voltage SourceMeter Instrument
12-96 2470-901-01 Rev. A / May 2019
Example
SOUR:FUNC VOLT
SOUR:VOLT:RANG 20
SENS:FUNC "CURR"
SENS:CURR:RANG 100e-6
SOUR:SWE:VOLT:LOG .1, 10, 20, 1e-3, 1, FIXED
Reset the instrument to its defaults.
Set the source function to voltage.
Set the source range to 20 V.
Set the measure function to current.
Set the current range to 100 μA.
Set up a log sweep that sweeps from 0.1 to 10 V in 20 steps with a source delay of 1 ms, a sweep count of 1,
and a fixed source range.
Also see
:INITiate[:IMMediate] (on page 12-143)
Sweep operation (on page 4-54)
STATus subsystem
The STATus subsystem controls the status registers of the instrument. For additional information on
the status model, see Status model (on page B-1
).
:STATus:CLEar
This function clears event registers.
Type Affected by Where saved Default value
Usage
:STATus:CLEar
Details
This command clears the event registers of the Questionable Event and Operation Event Register
set. It does not affect the Questionable Event Enable or Operation Event Enable registers.
Example
Clear the bits in the registers.
Also see
*CLS (on page A-2)