Section
14: TSP command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
14-96 2470-901-01 Rev. A / May 2019
node[N].execute()
This function starts test scripts on a remote TSP-Link node.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes (see Details)
Usage
node[N].execute("scriptCode")
N
The node number of this instrument (1 to 63)
scriptCode
A string containing the source code
Details
This command is only applicable to TSP-Link systems. You can use this command to use the remote
master node to run a script on the specified node. This function does not run test scripts on the
master node; only on the subordinate node when initiated by the master node.
This function may only be called when the group number of the node is different than the node of the
master.
This function does not wait for the script to finish execution.
Example 1
node[2].execute(sourcecode)
Runs script code on node 2. The code is in a string variable
called
sourcecode
.
Example 2
node[3].execute("x = 5")
Runs script code in string constant (
"x = 5"
) to set
x
equal
to 5 on node 3.
Example 3
node[32].execute(TestDut.source)
Runs the test script stored in the variable TestDut
(previously stored on the master node) on node 32.
Also see
tsplink.group (on page 14-269)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals