High Voltage SourceMeter Instrument Reference Manual Section 12:
2470-901-01 Rev. A / May 2019 12-57
[:SENSe[1]]:<function>:RELative:STATe
This command enables or disables the application of a relative offset value to the measurement.
Type Affected by Where saved Default value
Command and query Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
OFF (0)
Usage
[:SENSe[1]]:<function>:RELative:STATe <state>
[:SENSe[1]]:<function>:RELative:STATe?
ï‚§ Current: CURRent[:DC]
ï‚§ Resistance: RESistance
ï‚§ Voltage: VOLTage[:DC]
Disable the relative offset: OFF or 0
Enable the relative offset:
or
Details
When relative measurements are enabled, all subsequent measured readings are offset by the
relative offset value. You can enter a relative offset value or have the instrument acquire a relative
offset value.
Each returned measured relative reading is the result of the following calculation:
Displayed reading = Actual measured reading - Relative offset value
Example
:SENS:VOLT:REL 5
Set the measurement function to volts with a relative offset of
5 V and enable the relative offset function.
Also see
Relative offset (on page 4-46)
[:SENSe[1]]:<function>:RELative (on page 12-55)
[:SENSe[1]]:<function>:RELative:ACQuire (on page 12-56)