Section
4: Sourcing and measuring 2470 High Voltage SourceMeter Instrument
Reference Manual
4-62 2470-901-01 Rev. A / May 2019
For example sweeps, see Sweep programming examples (on page 4-63).
For detail on the commands and options listed above, see the following command descriptions:
• createconfigscript() (on page 14-46)
• smu.source.sweeplinear() (on page 14-168)
• smu.source.sweeplinearstep() (on page 14-170)
• smu.source.sweeplist() (on page 14-172)
• smu.source.sweeplog() (on page 14-174)
• smu.source.func (on page 14-160)
• smu.source.range (on page 14-166)
• smu.source.configlist.create() (on page 14-153)
• smu.source.configlist.store() (on page 14-158)
• smu.measure.func (on page 14-127)
• smu.measure.range (on page 14-141)
Aborting a sweep
Sweeps can be stopped for the following reasons:
• The limit set by the abort on limit setting was exceeded
• The trigger model is aborted
You can stop the sweep while it is in progress. When you stop the sweep, all sweep commands in the
trigger model are terminated.
Using the front panel:
Press the front-panel TRIGGER key for two seconds and select Abort Trigger Model.
Using SCPI commands:
Send the command:
:ABORt
Using TSP commands:
Send the command:
trigger.model.abort()

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals