12-132 2470-901-01 Rev. A / May 2019
:TRACe:SAVE:APPend
This command appends data from the reading buffer to a file on the USB flash drive.
Type Affected by Where saved Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:SAVE:APPend "<fileName>"
:TRACe:SAVE:APPend "<fileName>", "<bufferName>"
:TRACe:SAVE:APPend "<fileName>", "<bufferName>", <timeFormat>
:TRACe:SAVE:APPend "<fileName>", "<bufferName>", <timeFormat>, <start>, <end>
<fileName>
A string that indicates the name of the file on the USB flash drive in which to save the
reading buffer
<bufferName>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2) or the name of a user-defined buffer; if no buffer is specified,
defbuffer1
is used
<timeFormat>
Indicates how date and time information from the buffer is saved in the file on the
USB flash drive; the values are:
ï‚§ Dates, times, and fractional seconds are saved; the default value: FORMat
ï‚§ Relative timestamps are saved: RELative
ï‚§ Seconds and fractional seconds are saved: RAW
ï‚§ Timestamps are saved:
STAMp
<start>
Defines the starting point in the buffer to start saving data
<end>
Defines the ending point in the buffer to stop saving data
Details
If the file you specify does not exist on the USB flash drive, this command creates the file.
For options that save more than one item of time information, each item is comma-delimited. For
example, the default format is date, time, and fractional seconds for each reading.
The file extension .csv is appended to the filename if necessary. Any file extension other than .csv
generates an error.
The index column entry in the .csv file starts at 1 for each append operation.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals