High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-261
trigger.timer[N].wait()
This function waits for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
triggered = trigger.timer[N].wait(timeout)
triggered
Trigger detection indication
N
Trigger timer number (1 to 4)
timeout
Maximum amount of time in seconds to wait for the trigger
Details
If one or more trigger events were detected since the last time trigger.timer[N].wait() or
trigger.timer[N].clear() was called, this function returns immediately.
After waiting for a trigger with this function, the event detector is automatically reset and rearmed.
This is true regardless of the number of events detected.
Example
triggered = trigger.timer[3].wait(10)
print(triggered)
Waits up to 10 s for a trigger on timer 3.
If false is returned, no trigger was detected
during the 10 s timeout.
If
true
is returned, a trigger was detected.
Also see
trigger.timer[N].clear() (on page 14-251)
trigger.tsplinkin[N].clear()
This function clears the event detector for a LAN trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.tsplinkin[N].clear()
N
The trigger line (1 to 3) to clear

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals