2470
High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-193
trigger.blender[N].clear()
This function clears the blender event detector and resets the overrun indicator of blender N.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.blender[N].clear()
N
The blender number (up to two)
Details
This command sets the blender event detector to the undetected state and resets the overrun
indicator of the event detector.
Example
trigger.blender[2].clear()
Clears the event detector for blender 2.
Also see
None
trigger.blender[N].orenable
This attribute selects whether the blender performs OR operations or AND operations.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Trigger blender N reset
Configuration script
false (AND)
Usage
orenable = trigger.blender[N].orenable
trigger.blender[N].orenable = orenable
orenable
The type of operation:
ï‚§ true: OR operation
ï‚§
false
: AND operation
N
The blender number (up to two)
Details
This command selects whether the blender waits for any one event (OR) or waits for all selected
ev
ents (AND) before signaling an output event.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals