Section
14: TSP command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
14-36 2470-901-01 Rev. A / May 2019
bufferVar.startindex
This attribute indicates the starting index in a reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Restore configuration
Instrument reset
Power cycle
Not applicable
Not applicable
Usage
bufferVar.startindex = startIndex
startIndex
Starting index of the buffer
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer
Details
Use this attribute to find the starting index in a reading buffer.
Example
test1 = buffer.make(100)
smu.measure.count = 6
smu.measure.read(test1)
print(test1.startindex, test1.endindex, test1.capacity)
Create a buffer named test1 with a capacity of 100 readings.
Set the measure count to 6.
Make measurements and store them in buffer test1.
Get the start index, end index, and capacity of test1.
Output:
1, 6, 100
Also see
bufferVar.endindex (on page 14-21)
buffer.make() (on page 14-12)
Reading buffers (on page 6-1)
Remote buffer operation (on page 6-25)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals