Section
12: SCPI command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
12-62 2470-901-01 Rev. A / May 2019
When an index is deleted from a configuration list, the index numbers of the following indexes are
shifted up by one. For example, if you have a configuration list with 10 indexes and you delete index
3, the index that was numbered 4 becomes index 3, and the all the following indexes are renumbered
in sequence to index 9. Because of this, if you want to delete several nonconsecutive indexes in a
configuration list, it is best to delete the higher numbered index first, then the next lower index, and so
on. This also means that if you want to delete all the indexes in a configuration list, you must delete
index 1 repeatedly until all indexes have been removed.
Example
:SENSe:CONF:LIST:DELete "myMeasList"
Deletes a configuration list named
myMeasList
.
:SENSe:CONF:LIST:DELete "myMeasList", 2
Deletes configuration index 2 in a
configuration list named
myMeasList
.
Also see
Configuration lists (on page 4-83)
[:SENSe[1]]:CONFiguration:LIST:CREate (on page 12-61)
[:SENSe[1]]:CONFiguration:LIST:QUERy?
This command returns a list of TSP commands and parameter settings that are stored in the specified
configuration index.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
[:SENSe[1]]:CONFiguration:LIST:QUERy? "<name>", <index>
[:SENSe[1]]:CONFiguration:LIST:QUERy? "<name>", <index>, <fieldSeparator>
<name>
A string that represents the name of a measure configuration list
<index>
A number that defines a specific configuration index in the configuration list
<fieldSeparator>
A separator for the data:
ï‚§ Comma (default): COMMa
ï‚§ Semicolon: SEMicolon
ï‚§ New line: NEWLine
Details
This command recalls data for one configuration index.
For additional information about the information this command recalls when using a configuration list
query command, see Instrument settings stored in a measure configuration list (on page 4-86
).

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals