14: TSP command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
14-20 2470-901-01 Rev. A / May 2019
bufferVar.dates
This attribute contains the dates of readings that are stored in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Restore configuration
Instrument reset
Power cycle
Not applicable Not applicable
Usage
date = bufferVar.dates[N]
date
The date of readings stored in bufferVar element N
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2
) or a user-defined buffer
N
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in the
buffer
Details
The dates are formatted as month, day, year.
This is not available if the reading buffer style is set to compact.
Example
reset()
testData = buffer.make(50)
trigger.model.load("SimpleLoop", 3, 1, testData)
trigger.model.initiate()
waitcomplete()
print(testData.dates[1])
printbuffer(1, testData.n, testData.dates)
Create a reading buffer named testData,
configure the instrument to make three
measurements, and store the readings in
the buffer.
Print the first reading date.
Example output:
11/27/2017
Prints the dates for readings 1 through the
last reading in the buffer.
Example output:
11/27/2017, 11/27/2017,
11/27/2017
Also see
buffer.delete() (on page 14-9)
buffer.make() (on page 14-12)
bufferVar.clear() (on page 14-18)
print() (on page 14-99)
printbuffer() (on page 14-100)
Reading buffers (on page 6-1)
Remote buffer operation (on page 6-25)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
Model2470
TypeSourceMeter
CategoryMeasuring Instruments
Channels1
Voltage Source Range±1000 V
Current Source Range±1 A
Power100 W
Current Measurement Resolution10 pA
Measurement Resolution6½ digits
Display5 inch capacitive touchscreen
InterfaceUSB, GPIB, Ethernet
Supported Communication ProtocolsSCPI
Input/Output Connectors4-wire

Related product manuals