High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-37
bufferVar.statuses
This attribute contains the status values of readings in the reading buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R) Yes Restore configuration
Instrument reset
Power cycle
Not applicable Not applicable
Usage
statusInformation = bufferVar.statuses[N]
statusInformation
The status value when reading N of the specified buffer was acquired; refer to
Details
bufferVar
The name of the reading buffer, which may be a default buffer (defbuffer1 or
defbuffer2) or a user-defined buffer
N
The reading number N; can be any value from 1 to the number of readings in the
buffer; use the bufferVar.n command to determine the number of readings in
the buffer
Details
This command is not available if the buffer style is set to compact.
This read-only attribute is an array of status values for the readings in the buffer. The status values
are floating-point numbers that encode the status value. Refer to the following table for values.
Buffer status bits for sense measurements
Bit (hex) Name Decimal Description statusInformation parameter
0x0001
STAT_QUESTIONABLE
1
Measure status
questionable
buffer.STAT_QUESTIONABLE
0x0006
STAT_ORIGIN
6
A/D converter
from which
reading
originated; for
the 2470, this will
always be 0
(main)
buffer.STAT_ORIGIN
0x0008
STAT_TERMINAL
8
Measure
terminal, front is
1, rear is 0
buffer.STAT_TERMINAL
0x0010
STAT_LIMIT2_LOW
16
Measure status
limit 2 low
buffer.STAT_LIMIT2_LOW
0x0020
STAT_LIMIT2_HIGH
32
Measure status
limit 2 high
buffer.STAT_LIMIT2_HIGH
0x0040
STAT_LIMIT1_LOW
64
Measure status
limit 1 low
buffer.STAT_LIMIT1_LOW
0x0080
STAT_LIMIT1_HIGH
128
Measure status
limit 1 high
buffer.STAT_LIMIT1_HIGH
0x0100
STAT_START_GROUP
256
First reading in a
group
buffer.STAT_START_GROUP

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals