High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-179
Details
This command selects which set of input and output terminals the instrument uses. You can select
front panel or rear panel terminals.
If the output is on when you change from one set of terminals to the other, the output is turned off.
This command replaces the smu.measure.terminals command, which is deprecated.
Example
smu.terminals = smu.TERMINALS_FRONT
Use the front-panel terminals for measurements.
Also see
None
status.clear()
This function clears event registers.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
status.clear()
Details
This command clears the event registers of the Questionable Event and Operation Event Register
set. It does not affect the Questionable Event Enable or Operation Event Enable registers.
Example
status.clear()
Clear the bits in the registers
Also see
*CLS (on page A-2)
status.condition
This attribute stores the status byte condition register.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
statusByte = status.condition
statusByte
The status byte

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals