Section
12: SCPI command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
12-108 2470-901-01 Rev. A / May 2019
Example
SYST:ERR:NEXT?
Returns information on the next error in the event log. For example, if you sent
a command without a parameter, the return is:
-109,"Missing parameter;1;2017/05/06 12:57:04.484"
Also see
:SYSTem:EVENtlog:NEXT? (on page 12-110)
:SYSTem:ERRor:CODE[:NEXT]?
This command reads the oldest error code.
Type Affected by Where saved Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:SYSTem:ERRor:CODE[:NEXT]?
Details
This command returns the numeric code of the next error in the event log. The error is cleared from
the queue after being read.
This command returns only error messages from the event log. To return information and warning
messages, see :SYSTem:EVENtlog:NEXT?
Example
SYST:ERR:CODE?
Returns the error code of the next error in the event log.
For example, if error -222, Parameter data out of range error,
occurred, the output is:
-222
Also see
:SYSTem:EVENtlog:NEXT? (on page 12-110)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals