Page #718 background imageLoading...
Page #718 background image
Section
14: TSP command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
14-78 2470-901-01 Rev. A / May 2019
file.write(fileNumber, "Tested to Company Standard ABC.123\n")
-- Ensure a hurry-up of data written to the file before close or script end
file.flush(fileNumber)
-- Close the data file
file.close(fileNumber)
end
This example writes a string that indicates that the readings were made for a certain reason, such as to test to
a company standard.
Also see
None
file.mkdir()
This function creates a directory at the specified path on the USB flash drive.
Type TSP-Link accessible
Affected by Where saved Default value
Function No
Usage
file.mkdir("path")
path
A string that contains the path of the directory
Details
The directory path must be absolute. The name of the directory must not already exist on the flash
drive.
Example
file.mkdir("TestData")
Create a new directory named TestData.
Also see
None

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals