High Voltage SourceMeter Instrument Reference Manual Section 4:
Sourcing and measuring
2470-901-01 Rev. A / May 2019 4-43
Setting autozero
You can enable or disable automatic referencing. You can also request a one-time refresh of the
reference values.
The reference setting is stored with the selected measure function.
To set autozero using the front panel:
1. Press the FUNCTION key.
2. Select the source and measure combination.
3. Press the MENU key.
4. Under Measure, select Settings.
5. For Auto Zero, select On or Off.
6. If Off is selected, you can select the Once option to send a one-time refresh.
7. Select HOME to return to the operating display.
To set autozero using SCPI commands:
Refer to the following commands:
• [:SENSe[1]]:<function>:AZERo[:STATe] (on page 12-46)
• [:SENSe[1]]:AZERo:ONCE (on page 12-59)
To set autozero using TSP commands:
Refer to the following commands:
• smu.measure.autozero.enable (on page 14-113)
• smu.measure.autozero.once() (on page 14-114)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals