Section
14: TSP command reference 2470 High Voltage SourceMeter Instrument
Reference Manual
14-206 2470-901-01 Rev. A / May 2019
trigger.lanin[N].clear()
This function clears the event detector for a LAN trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.lanin[N].clear()
N
The LAN event number (1 to 8) to clear
Details
The trigger event detector enters the detected state when an event is detected. This function clears a
trigger event detector and discards the history of the trigger packet.
This function clears all overruns associated with this LAN trigger.
Example
trigger.lanin[5].clear()
Clears the event detector with LAN event trigger 5.
Also see
trigger.lanin[N].overrun (on page 14-207)
trigger.lanin[N].edge
This attribute sets the trigger operation and detection mode of the specified LAN event.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Restore configuration
Instrument reset
Power cycle
Configuration script trigger.EDGE_EITHER
Usage
edgeMode = trigger.lanin[N].edge
trigger.lanin[N].edge = edgeMode
edgeMode
The trigger mode; see the Details for more information
N
The LAN event number (1 to 8)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals