14: TSP command reference 2470 High Voltage SourceMeter Instrument
14-284 2470-901-01 Rev. A / May 2019
For the termSequence parameter, use the same values listed in the table above for type. There are
four possible combinations, all of which are made up of line feeds (LF or 0x10) and carriage returns
(CR or 0x13). For TSP-enabled devices, the default is tspnet.TERM_LF. For devices that are not
TSP-enabled, the default is tspnet.TERM_CRLF.
Example
deviceID = tspnet.connect("192.0.2.1")
if deviceID then
tspnet.termination(deviceID, tspnet.TERM_LF)
Sets termination type for IP address
192.0.2.1 to TERM_LF.
Also see
tspnet.connect() (on page 14-277)
tspnet.disconnect() (on page 14-278)
tspnet.timeout
This attribute sets the timeout value for the tspnet.connect(), tspnet.execute(), and tspnet.read()
commands.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) No Restore configuration
Instrument reset
Configuration script 20.0 (20 s)
Usage
value = tspnet.timeout
tspnet.timeout = value
The timeout duration in seconds (1 ms to 30.0 s)
Details
This attribute sets the amount of time the tspnet.connect(), tspnet.execute(), and
tspnet.read()
commands will wait for a response.
The time is specified in seconds. The timeout may be specified to millisecond resolution, but is only
accurate to the nearest 10 ms.
Example
Sets the timeout duration to 2 s.
Also see
tspnet.connect() (on page 14-277)
tspnet.execute() (on page 14-279)
tspnet.read() (on page 14-281)