12: SCPI command reference  2470 High Voltage SourceMeter Instrument 
 
12-66 2470-901-01 Rev. A / May 2019 
[:SENSe[1]]:COUNt 
This command sets the number of measurements to make when a measurement is requested. 
Type  Affected by  Where saved  Default value 
Command and query  Recall settings 
Instrument reset 
Save settings  1 
Usage 
[:SENSe[1]]:COUNt <n> 
[:SENSe[1]]:COUNt <DEF|MIN|MAX> 
[:SENSe[1]]:COUNt? 
[:SENSe[1]]:COUNt? <DEF|MIN|MAX> 
 
The number of measurements (1 to 300,000) 
The DEFault, MINimum, or MAXimum value
Details 
This command sets the number of measurements that are made when a measurement is requested. 
This command does not affect the trigger model. 
This command sets the count for all measure functions. 
If you set the count to a value that is larger than the capacity of the reading buffer and the buffer fill 
mode is set to continuous, the buffer wraps until the number of readings specified have occurred. The 
earliest readings in the count are overwritten. If the buffer is set to fill once, readings stop when the 
buffer is filled, even if the count is not complete. 
To get better performance from the instrument, use the Simple Loop trigger model template instead 
of using the count command. 
Example 
 
:TRAC:CLEAR 
:COUN 10 
:MEAS? 
Clear data from the reading buffer. 
Set the count to 10. 
Make ten measurements. 
Returns the last measurement. 
Example output: 
-5.693831E-05
Read all ten measurements. 
Example output: 
-7.681046E-05,-2.200288E-04,-9.086048E-05,-6.388056E-05,-7.212282E-05,-4.874761E-0
5,-4.741654E-04,-6.811028E-05,-5.110232E-05,-5.693831E-05