High Voltage SourceMeter Instrument Reference Manual Section 12:
2470-901-01 Rev. A / May 2019 12-43
Example 2
RES:AVER:TCON MOV
For resistance measurements, set the averaging filter type to
moving average, with a filter count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:TCON MOV
For voltage measurements, set the averaging filter type to
moving average, with a filter count of 10.
Enable the averaging filter.
Also see
Filtering measurement data (on page 5-25)
[:SENSe[1]]:<function>:AVERage[:STATe] (on page 12-43)
[:SENSe[1]]:<function>:AVERage:TCONtrol (on page 12-44)
[:SENSe[1]]:<function>:AVERage[:STATe]
This command enables or disables the averaging filter for measurements of the selected function.
Type Affected by Where saved Default value
Instrument reset
Power cycle
Measure configuration list
Measure configuration list
Usage
[:SENSe[1]]:<function>:AVERage[:STATe] <state>
[:SENSe[1]]:<function>:AVERage[:STATe]?
ï‚§ Current: CURRent[:DC]
ï‚§ Resistance: RESistance
ï‚§ Voltage:
The filter status; set to one of the following values:
ï‚§ Disable the averaging filter: OFF or 0
ï‚§ Enable the averaging filter:
or
Details
This command enables or disables the averaging filter. When this is enabled, the reading returned by
the instrument is an averaged value, taken from multiple measurements. The settings of the filter
count and filter type for the selected measure function determines how the reading is averaged.
If you send this command without the <function> parameter, it sets the state of the averaging filter
for all functions.