2470
High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-279
tspnet.execute()
This function sends a command string to the remote device.
Type TSP-Link accessible
Affected by Where saved Default value
Function
No
Usage
tspnet.execute("connectionID", "commandString")
value1 = tspnet.execute("connectionID", "commandString", formatString)
value1, value2 = tspnet.execute("connectionID", "commandString", formatString)
value1, ..., valueN = tspnet.execute("connectionID", "commandString", formatString)
connectionID
The connection ID returned from tspnet.connect()
commandString
The command to send to the remote device
value1
The first value decoded from the response message
value2
The second value decoded from the response message
valueN
The
N
th value decoded from the response message; there is one return value for
each format specifier in the format string
...
One or more values separated with commas
formatString
Format string for the output
Details
This command sends a command string to the remote instrument. A termination is added to the
command string when it is sent to the remote instrument (tspnet.termination()). You can als
o
s
pecify a format string, which causes the command to wait for a response from the remote
instrument. The 2470 decodes the response message according to the format specified in the format
string and returns the message as return values from the function (see tspnet.read() for format
specifiers).
When this command is sent to a TSP-enabled instrument, the 2470 suspends operation until a
timeout error is generated or until the instrument responds. The TSP prompt from the remote
instrument is read and discarded. The 2470 places any remotely generated errors and events into its
event queue.When the optional format string is not specified, this command is equivalent to
tspnet.write(), except that a termination is automatically added to the end of the command.
Example 1
tspnet.execute(instrumentID, "runScript()")
Command the remote device to run a script named runScript.
Example 2
tspnet.timeout = 5
id_instr = tspnet.connect("192.0.2.23", 23, "*rst\r\n")
tspnet.termination(id_instr, tspnet.TERM_CRLF)
tspnet.execute(id_instr, "*idn?")
print("tspnet.execute returns:", tspnet.read(id_instr))
Print the *idn? string from the remote device.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals