14: TSP command reference 2470 High Voltage SourceMeter Instrument
14-292 2470-901-01 Rev. A / May 2019
Example
userstring.add("assetnumber", "236")
value = userstring.get("assetnumber")
print(value)
Create the user-defined string
assetnumber, set to a value of 236.
Read the value associated with the
user-defined string named assetnumber.
Store it in a variable called value, then print
the variable value.
Output:
Also see
userstring.add() (on page 14-289)
userstring.catalog() (on page 14-290)
userstring.delete() (on page 14-291)
waitcomplete()
This function waits for all previously started overlapped commands to complete.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
waitcomplete()
waitcomplete(group)
Specifies which TSP-Link group on which to wait
Details
There are two types of instrument commands:
• Overlapped commands: Commands that allow the execution of subsequent commands while
i
nstrument operations of the overlapped command are still in progress.
• Sequential commands: Commands whose operations must finish before the next command is
executed.
The waitcomplete() command suspends the execution of commands until the instrument
operations of all previous overlapped commands are finished. This command is not needed for
sequential commands.
A group number may only be specified when this node is the master node.
If no group is specified, the local group is used.
If zero (0) is specified for the group, this function waits for all nodes in the system.
Any nodes that are not assigned to a group (group number is 0) are part of the master node's group.