12: SCPI command reference 2470 High Voltage SourceMeter Instrument
12-44 2470-901-01 Rev. A / May 2019
Example 1
CURR:AVER:TCON MOV
Set the averaging filter type to moving average, with a filter
count of 10.
Enable the averaging filter.
Example 2
RES:AVER:TCON MOV
Set the averaging filter type to moving average, with a filter
count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:TCON MOV
Set the averaging filter type to moving average, with a filter
count of 10.
Enable the averaging filter.
Also see
Filtering measurement data (on page 5-25)
[:SENSe[1]]:<function>:AVERage:COUNt (on page 12-42)
[:SENSe[1]]:<function>:AVERage:TCONtrol (on page 12-44)
[:SENSe[1]]:<function>:AVERage:TCONtrol
This command sets the type of averaging filter that is used for the selected measure function when the
measurement filter is enabled.
Type Affected by Where saved Default value
Command and query Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
REP
Usage
[:SENSe[1]]:<function>:AVERage:TCONtrol <type>
[:SENSe[1]]:<function>:AVERage:TCONtrol?
ï‚§ Current: CURRent[:DC]
ï‚§ Resistance: RESistance
ï‚§ Voltage:
The filter type to use when filtering is enabled; set to one of the following values:
ï‚§ Repeating filter: REPeat
ï‚§ Moving filter: