:STATus:OPERation:CONDition?
This command reads the Operation Event Register of the status model.
Type Affected by Where saved Default value
Usage
:STATus:OPERation:CONDition?
Details
This command reads the contents of the Operation Condition Register, which is one of the Operation
Event Registers.
For detail on interpreting the value of a register, see Understanding bit settings (on page B-15
).
Example
Returns the contents of the Operation Condition Register.
Also see
Operation Event Register (on page B-7)
:STATus:OPERation:ENABle
This command sets or reads the contents of the Operation Event Enable Register of the status model.
Type Affected by Where saved Default value
Usage
:STATus:OPERation:ENABle <n>
:STATus:OPERation:ENABle?
The status of the operation status register
Details
This command sets or reads the contents of the Enable register of the Operation Event Register.
When one of these bits is set, when the corresponding bit in the Operation Event Register or
Operation Condition Register is set, the OSB bit in the Status Byte Register is set.
When sending binary values, preface <n> with #b. When sending hexadecimal values, preface <n>
with #h. No preface is needed when sending decimal values.