High Voltage SourceMeter Instrument Reference Manual Section 12:
2470-901-01 Rev. A / May 2019 12-65
[:SENSe[1]]:CONFiguration:LIST:STORe
This command stores the active measure settings into the named configuration list.
Type Affected by Where saved Default value
Instrument reset
Usage
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>"
[:SENSe[1]]:CONFiguration:LIST:STORe "<name>", <index>
A string that represents the name of a measure configuration list
A number that defines a specific configuration index in the configuration list
Details
Use this command to store the active measure settings to a configuration index in a configuration list.
If the index parameter is not provided, the new settings are appended to the end of the list. The index
only stores the active settings for a single active measure function.
Configuration lists are not saved when the instrument is turned off or reset. To save a configuration
list, create a configuration script to save instrument settings, including any defined configuration lists.
Refer to Instrument settings stored in a measure configuration list (on page 4-86
) for a complete list of
measure settings that the instrument stores.
Example
:SENSe:CONF:LIST:STOR "MyConfigList"
Stores the active settings of the instrument to
the end of the configuration list named
:SENSe:CONF:LIST:STOR "MyConfigList", 5
Stores the active settings of the instrument to
the configuration list named MyConfigList
in configuration index 5.
Also see
Configuration lists (on page 4-83)
[:SENSe[1]]:CONFiguration:LIST:CREate (on page 12-61)
*SAV (on page 12-9)