12: SCPI command reference 2470 High Voltage SourceMeter Instrument
12-100 2470-901-01 Rev. A / May 2019
:STATus:PRESet
This command resets all bits in the status model.
Type Affected by Where saved Default value
Usage
:STATus:PRESet
Details
This function clears the event registers and the enable registers for operation and questionable. It will
not clear the Service Request Enable Register (*SRE) to Standard Request Enable Register (*ESE).
Preset does not affect the event queue.
The Standard Event Status Register is not affected by this command.
Example
Also see
Status model (on page B-1)
:STATus:QUEStionable:CONDition?
This command reads the Questionable Condition Register of the status model.
Type Affected by Where saved Default value
Query only Not applicable Not applicable Not applicable
Usage
:STATus:QUEStionable:CONDition?
Details
This command reads the contents of the Questionable Condition Register, which is one of the
Questionable Event Registers.
For detail on interpreting the value of a register, see Understanding bit settings (on page B-15
).
Example
Reads the Questionable Condition Register.
Also see
Questionable Event Register (on page B-6)
Understanding bit settings (on page B-15)