Appendix
A: Common commands 2470 High Voltage SourceMeter Instrument
Reference Manual
A-6 2470-901-01 Rev. A / May 2019
*LANG
This command determines which command set is used by the instrument.
Type Affected by Where saved Default value
Command and query Not applicable Nonvolatile memory SCPI
Usage
*LANG <commandSet>
*LANG?
<commandSet>
The command set to be used:
ï‚§ TSP
ï‚§
SCPI
Details
The remote command sets that are available include:
• SCPI: An instrument-specific language built on the SCPI standard.
• TSP: A scripting programming language that contains instrument-specific control commands that
can be executed from a stand-alone instrument. You can use TSP to send individual commands
or use it to combine commands into scripts.
If you change the command set, reboot the instrument.
You cannot combine the command sets.
Example
*LANG TSP
*LANG?
Set the command set to TSP.
Verify setting by sending the command set query.
Output:
TSP
The TSP command set is in use.
Also see
Status model (on page B-1)
*OPC
This command sets the operation complete (OPC) bit after all pending commands, including overlapped
commands, have been executed.
Type Affected by Where saved Default value
Command and query
Not applicable
Not applicable
Not applicable
Usage
*OPC
*OPC?

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals