14: TSP command reference 2470 High Voltage SourceMeter Instrument
14-288 2470-901-01 Rev. A / May 2019
Details
The tspnet.write() function sends inputString to the remote instrument. It does not wait for
command completion on the remote instrument.
The 2470 sends inputString to the remote instrument exactly as indicated. The inputString
must contain any necessary new lines, termination, or other syntax elements needed to complete
properly.
Because tspnet.write() does not process output from the remote instrument, do not send
commands that generate too much output without processing the output. This command can stop
executing if there is too much unprocessed output from previous commands.
Example
tspnet.write(myID, "runscript()\r\n")
Commands the remote instrument to execute
a command or script named runscript()
on a remote device identified in the system
as
.
Also see
tspnet.connect() (on page 14-277)
tspnet.read() (on page 14-281)
upgrade.previous()
This function returns to a previous version of the 2470 firmware.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
upgrade.previous()
Details
This function allows you to revert to an earlier version of the firmware.
When you send this function, the instrument searches the USB flash drive in the front-panel USB port
for an upgrade file. If the file is found, the instrument performs the upgrade. An error is returned if an
upgrade file is not found.
Use this command with caution. Make sure your instrument can support the earlier version and that
there are no compatibility issues. Check with Keithley Instruments before using this command if you
have questions.
Also see
Upgrading the firmware (on page 10-3)
upgrade.unit() (on page 14-289)