12: SCPI command reference 2470 High Voltage SourceMeter Instrument
12-114 2470-901-01 Rev. A / May 2019
:SYSTem:LFRequency?
This query contains the power line frequency setting that is used for NPLC calculations.
Type Affected by Where saved Default value
Usage
:SYSTem:LFRequency?
Details
The instrument automatically detects the power line frequency when the instrument is powered on.
Power line frequency can be 50 Hz or 60 Hz.
Example
Check the line frequency.
Also see
None
:SYSTem:PASSword:NEW
This command stores the instrument password.
Type Affected by Where saved Default value
Usage
:SYSTem:PASSword:NEW "<password>"
A string that contains the instrument password (maximum 30 characters)
Details
When the access to the instrument is set to protected or lockout, this is the password that is used to
gain access.
If you forget the password, you can reset the password to the default:
1. On the front panel, press MENU.
2. Under System, select Info/Manage.
3. Select Password Reset.
You can also reset the password and the LAN settings from the rear panel by inserting a straightened
paper clip into hole below LAN RESET.
Example
SYST:PASS:NEW "N3wpa55w0rd"
Change the password of the instrument to N3wpa55w0rd.
Also see
:SYSTem:ACCess (on page 12-103)