2470
High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-209
Details
Generates and sends a LAN trigger packet for the LAN event number specified.
Sets the pseudo line state to the appropriate state.
The following indexes provide the listed LXI events:
ï‚§ 1:LAN0
ï‚§ 2:LAN1
ï‚§ 3:LAN2
 …
ï‚§ 8:LAN7
Example
trigger.lanout[5].assert()
Creates a trigger with LAN trigger 5.
Also see
lan.lxidomain (on page 14-87)
trigger.lanin[N].clear() (on page 14-206)
trigger.lanin[N].overrun (on page 14-207)
trigger.lanin[N].wait() (on page 14-208)
trigger.lanout[N].assert() (on page 14-208)
trigger.lanout[N].ipaddress (on page 14-211)
trigger.lanout[N].protocol (on page 14-212)
trigger.lanout[N].stimulus (on page 14-213)
trigger.lanout[N].connect()
This function prepares the event generator for outgoing trigger events.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
trigger.lanout[N].connect()
N
The LAN event number (1 to 8)
Details
This command prepares the event generator to send event messages. For TCP connections, this
opens the TCP connection.
The event generator automatically disconnects when either the protocol or IP address for this event is
changed.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals