14: TSP command reference 2470 High Voltage SourceMeter Instrument
14-208 2470-901-01 Rev. A / May 2019
Also see
trigger.lanin[N].clear() (on page 14-206)
trigger.lanin[N].wait() (on page 14-208)
trigger.lanout[N].assert() (on page 14-208)
trigger.lanout[N].stimulus (on page 14-213)
trigger.lanin[N].wait()
This function waits for an input trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
triggered = trigger.lanin[N].wait(timeout)
Trigger detection indication (true or false)
The trigger packet over LAN to wait for (1 to 8)
Maximum amount of time in seconds to wait for the trigger event
Details
If one or more trigger events have been detected since the last time trigger.lanin[N].wait()
or trigger.lanin[N].clear() was called, this function returns immediately.
After waiting for a LAN trigger event with this function, the event detector is automatically reset and
rearmed regardless of the number of events detected.
Example
triggered = trigger.lanin[5].wait(3)
Wait for a trigger event with LAN trigger 5
with a timeout of 3 s.
Also see
trigger.lanin[N].clear() (on page 14-206)
trigger.lanin[N].overrun (on page 14-207)
trigger.lanout[N].assert() (on page 14-208)
trigger.lanout[N].stimulus (on page 14-213)
trigger.lanout[N].assert()
This function simulates the occurrence of the trigger and generates the corresponding event.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.lanout[N].assert()
The LAN event number (1 to 8)