High Voltage SourceMeter Instrument Reference Manual Section 12:
2470-901-01 Rev. A / May 2019 12-67
Also see
:MEASure? (on page 12-4)
:TRACe:DATA? (on page 12-120)
:TRIGger:LOAD "SimpleLoop" (on page 12-188)
[:SENSe[1]]:FUNCtion[:ON]
This command selects the active measure function.
Type Affected by Where saved Default value
Command and query Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
CURR
Usage
[:SENSe[1]]:FUNCtion[:ON] "<function>"
[:SENSe[1]]:FUNCtion[:ON]?
A string that contains the measure function:
ï‚§ Current: CURRent[:DC]
ï‚§ Resistance: RESistance
ï‚§ Voltage:
Details
Set this command to the type of measurement you want to make.
Reading this command returns the measure function that is presently active.
Example
Make the voltage measurement function the active function.
Also see
Making resistance measurements (on page 4-25)
Source and measure using SCPI commands (on page 4-31)