High Voltage SourceMeter Instrument Reference Manual Section 12:
2470-901-01 Rev. A / May 2019 12-101
:STATus:QUEStionable:ENABle
This command sets or reads the contents of the questionable event enable register of the status model.
Type Affected by Where saved Default value
Usage
:STATus:QUEStionable:ENABle <n>
:STATus:QUEStionable:ENABle?
The value of the register (0 to 65535)
Details
This command sets or reads the contents of the Enable register of the Questionable Event Register.
When one of these bits is set, when the corresponding bit in the Questionable Event Register or
Questionable Condition Register is set, the MSB and QSM bits in the Status Byte Register are set.
For detail on interpreting the value of a register, see Understanding bit settings (on page B-15
).
Example
:STAT:QUES:ENAB 8
:STAT:QUES:ENAB?
Enable bit 4, Limit 3 Fail, when the limit test 3 failure value is
exceeded. Check to see that the value was set.
Also see
Questionable Event Register (on page B-6)
:STATus:QUEStionable:MAP
This command queries mapped event numbers or maps event numbers to bits in the event registers.
Type Affected by Where saved Default value
Usage
:STATus:QUEStionable:MAP <bitNumber>, <setEvent>
:STATus:QUEStionable:MAP <bitNumber>, <setEvent>, <clearEvent>
:STATus:QUEStionable:MAP? <bitNumber>
The bit number that is mapped to an event (0 to 14)
The number of the event that sets the bits in the condition and event registers; 0 if no
mapping
The number of the event that clears the bit in the condition register; 0 if no mapping