EasyManuals Logo

Tektronix Keithley SourceMeter 2470 Reference Manual

Tektronix Keithley SourceMeter 2470
995 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #979 background imageLoading...
Page #979 background image
2470
High Voltage SourceMeter Instrument Reference Manual Appendix B:
Status model
2470-901-01 Rev. A / May 2019 B-17
When using the SCPI interface, this command does not affect the Questionable Event Enable
Register or Operation Event Enable Register sets.
Status model programming examples
The following examples illustrate how to generate an SRQ using the status model.
SRQ when the SMU reaches its source limit
This example demonstrates how to generate a service request (SRQ) when the source-measure unit
(SMU) detects it has reached its source limit. After configuring the status model, in this example the
source will be configured to output current and measure voltage. If the output terminals of the SMU
are left open when running this example, the SMU will reach its source limit and generate an SRQ.
Using SCPI commands:
*RST
STAT:CLE
STAT:OPER:MAP 0, 5080, 5081
STAT:OPER:ENAB 1
*SRE 128
SOUR:FUNC CURR
SOUR:CURR:RANG 1e-3
SOUR:CURR 1e-3
SOUR:CURR:VLIM 1
SENS:FUNC "VOLT"
OUTP ON
READ?
OUTP OFF

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals