Section
15: Frequently asked questions 2470 High Voltage SourceMeter Instrument
Reference Manual
15-6 2470-901-01 Rev. A / May 2019
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. Select the appropriate Command Set.
You are prompted to confirm the change to the command set and reboot.
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the SCPI command set from a remote interface:
Send the command:
*LANG SCPI
Reboot the instrument.
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
How do I upgrade the firmware?
Do not turn off power or remove the USB flash drive until the upgrade process is complete.
Do not initialize TSP-Link before starting the upgrade.
You can upgrade or downgrade the firmware from the front panel or from the virtual front panel.
Refer to Using the 2470 virtual front panel (on page 2-33
) for information.

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals