2470
High Voltage SourceMeter Instrument Reference Manual Section 14:
TSP command reference
2470-901-01 Rev. A / May 2019 14-185
status.questionable.condition
This attribute reads the Questionable Condition Register of the status model.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
questionableRegister = status.questionable.condition
questionableRegister
The value of the register (0 to 65535)
Details
This command reads the contents of the Questionable Condition Register, which is one of the
Questionable Event Registers.
For detail on interpreting the value of a register, see Understanding bit settings (on page B-15
).
Example
print(status.questionable.condition)
Reads the Questionable Condition Register.
Also see
Questionable Event Register (on page B-6)
Understanding bit settings (on page B-15)
status.questionable.enable
This attribute sets or reads the contents of the questionable event enable register of the status model.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
status.preset()
Not applicable
0
Usage
questionableRegister = status.questionable.enable
status.questionable.enable = questionableRegister
questionableRegister
The value of the register (0 to 65535)
Details
This command sets or reads the contents of the Enable register of the Questionable Event Register.
When one of these bits is set, when the corresponding bit in the Questionable Event Register or
Questionable Condition Register is set, the MSB and QSM bits in the Status Byte Register are set.
For detail on interpreting the value of a register, see Understanding bit settings (on page B-15
).
Example
status.questionable.enable = 17
print(status.questionable.enable)
Set bits 0 and 4 of the Questionable Event Enable Register.
Returns 17, which indicates the register was set correctly.
Also see
Questionable Event Register (on page B-6)

Table of Contents

Other manuals for Tektronix Keithley SourceMeter 2470

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley SourceMeter 2470 and is the answer not in the manual?

Tektronix Keithley SourceMeter 2470 Specifications

General IconGeneral
BrandTektronix
ModelKeithley SourceMeter 2470
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals