14: TSP command reference 2470 High Voltage SourceMeter Instrument Referen
14-278 2470-901-01 Rev. A / May 2019
You can simultaneously connect to a maximum of 32 remote devices.
Example 1
instrumentID = tspnet.connect("192.0.2.1")
if instrumentID then
-- Use instrumentID as needed here
tspnet.disconnect(instrumentID)
Connect to a TSP-enabled device.
Example 2
instrumentID = tspnet.connect("192.0.2.1", 1394, "*rst\r\n")
if instrumentID then
-- Use instrumentID as needed here
tspnet.disconnect(instrumentID)
Connect to a device that is not TSP-enabled.
Also see
localnode.prompts (on page 14-91)
localnode.showevents (on page 14-94)
tspnet.tsp.abortonconnect (on page 14-285)
tspnet.disconnect() (on page 14-278)
tspnet.disconnect()
This function disconnects a specified TSP-Net session.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
tspnet.disconnect(connectionID)
The connection ID returned from tspnet.connect()
Details
This function disconnects the two devices by closing the connection. The connectionID is the
session handle returned by tspnet.connect().
For TSP-enabled devices, this aborts any remotely running commands or scripts.
Example
testID = tspnet.connect("192.0.2.0")
-- Use the connection
tspnet.disconnect(testID)
Create a TSP-Net session.
Close the session.
Also see
tspnet.connect() (on page 14-277)